NABH CSSD checklist: what assessors ask to see

A practical preparation checklist for CSSD layout, cleaning, packing, sterilization monitoring, storage, recall, audits and training records.

Assessors visiting a CSSD ask the same question in many ways: can you prove that every item used on a patient was cleaned, packed and sterilized correctly — and find it again if something went wrong? This checklist groups what teams are usually asked to show. It is a preparation aid, not the standard itself: always work from the current NABH standards for your hospital type and your own infection-control policy.

1. Layout and workflow

  • One-way flow from dirty (decontamination) to clean (inspection and packing) to sterile (storage and issue), with physical separation.
  • Separate receiving and issue counters, or separate times if space is shared.
  • Written SOPs for each step, approved and within their review date, and evidence that staff have read them.

2. Cleaning and decontamination

  • Instrument count checked at receiving, with missing or damaged items recorded.
  • Washer-disinfector or ultrasonic loads recorded: programme, temperature, result and operator.
  • Inspection for cleanliness and function before packing.

3. Packing and labelling

  • Each pack labelled with a unique number, contents, packer, date and expiry (or event-related shelf-life statement).
  • A second check for critical sets where your policy asks for it.
  • Internal chemical indicators inside packs where required.

4. Sterilization monitoring

  • Bowie-Dick test daily, before the first load, for each pre-vacuum steam sterilizer — with the result filed.
  • Physical parameters (time, temperature, pressure) recorded for every load, ideally from the machine printout or data port.
  • Chemical indicators per pack and per load.
  • Biological indicators at the frequency in your policy, and for every load containing implants — with implant loads held until the result is known.
  • Release of each load signed by an authorised person.

5. Storage and issue

  • Sterile store temperature and humidity logged, with action taken when limits are crossed.
  • First-expiry-first-out; expired packs withdrawn and reprocessed, with a discard register.
  • Issue records showing which pack went to which department, and use records linking packs to patients.

6. Recall, audit and training

  • A written recall procedure, and records of each recall from start to closure.
  • Periodic CSSD audits with corrective and preventive actions (CAPA) and their verification.
  • Training and competency records for every CSSD staff member.

How software helps

With CareSterile, most of these records are produced as a side effect of scanning: every register prints in A4 format, every pack is traceable from the patient back to the load, and a recall lists affected packs and patients in one search. See NABH & JCI compliance.


More guides: The CSSD workflow, step by step · Sterilization indicators explained: Bowie-Dick, chemical and biological · Barcode vs RFID vs 2D codes for instrument tracking · How to run a CSSD recall in minutes, not days · CSSD KPIs: 12 indicators worth tracking, with formulas · Loaner set management: a safe process from booking to return · Sterile storage and shelf life: time-based vs event-related · The Bowie-Dick test: how to run it, read it and record it · Buying CSSD software: 8 questions to ask every vendor ·

See your own sets tracked end to end

30-minute online demo with the Caresoft team. Bring your set list — we will show it in the system.

Book a demo