Sterilization indicators explained: Bowie-Dick, chemical and biological

What each indicator class proves, when to use it, and how to record results so they hold up in an audit.

Three kinds of evidence

No single test proves a load is sterile. CSSDs combine physical monitoring (the sterilizer's own readings), chemical indicators and biological indicators. Each answers a different question.

IndicatorWhat it showsTypical use
Bowie-Dick testAir removal and steam penetration in a pre-vacuum steam sterilizerDaily, in an empty chamber, before the first load
Chemical — Class 1 (process)The pack went through a process (e.g. autoclave tape)Outside every pack
Chemical — Class 4 / 5 / 6One or more critical variables were reached inside the packInside packs; Class 5 integrators often used for release
Biological indicator (BI)Highly resistant spores were killedRoutine monitoring as per policy, and every load with implants

Chemical indicator classes follow ISO 11140-1. Steam and hydrogen peroxide BIs usually use Geobacillus stearothermophilus spores; ETO and dry heat BIs usually use Bacillus atrophaeus. Read-out times depend on the product — follow the manufacturer's instructions.

Recording results so they hold up

  • Record indicator type, lot number and expiry, not only "pass".
  • Link each result to the sterilizer, load number and date.
  • Record who read the result and when.
  • For a failed Bowie-Dick, stop using the sterilizer until it is fixed and retested.
  • For a positive BI, quarantine the load and start a recall for anything already issued.

How CareSterile handles this

Each indicator is logged against its load. A steam sterilizer is blocked for the day until its Bowie-Dick test passes, a load cannot be released until its indicators pass, and a positive BI opens a recall listing every affected pack and patient.


More guides: The CSSD workflow, step by step · Barcode vs RFID vs 2D codes for instrument tracking · How to run a CSSD recall in minutes, not days · NABH CSSD checklist: what assessors ask to see · CSSD KPIs: 12 indicators worth tracking, with formulas · Loaner set management: a safe process from booking to return · Sterile storage and shelf life: time-based vs event-related · The Bowie-Dick test: how to run it, read it and record it · Buying CSSD software: 8 questions to ask every vendor ·